X-ray Dosimetry Tools

Tube output, kVp/mAs adjustment, half-value layer and inverse-square law for medical physicists

kVp / mAs adjustment (15 percent rule, mAs square law)

Required new mAs
—
mAs

15% rule: 15% kVp increase ≈ doubling mAs for image density. Entrance dose ∝ mAs × kVp².

Half-Value Layer (HVL) from two attenuation measurements

HVL
—
mm Al
Linear attenuation coefficient µ
—
cm⁻¹

HVL = ln(2) / µ. µ derived by linear regression of ln(I) vs thickness from the two filter readings.

Inverse-square law (dose at new distance)

Dose at new distance
—
mSv

I₂ = I₁ × (d₁ / d₂)². Doubling distance → ¼ dose. Useful for radiation safety planning at new working distances.

Tube output: dose ↔ mAs

Entrance skin dose (ESD)
—
µGy
Air kerma free-in-air at SSD
—
µGy

K_air = Y(kVp) × mAs × (100 / SSD)² · (kVp/kVp_ref)². ESD = K_air × BSF.

Sources

  • IAEA Human Health Series No. 4 (2007). Comprehensive Clinical Audits of Diagnostic Radiology Practices.
  • Bushberg JT et al. The Essential Physics of Medical Imaging, 4th ed. (2020). Chapters 6–7 (X-ray production, dosimetry).
  • IPEM Report 78 (2002). Catalogue of Diagnostic X-ray Spectra and Other Data.
  • NCRP Report No. 168 (2010) — fluoroscopy dose management.
  • 15% rule (kVp/mAs trade-off) — Carlton & Adler, Principles of Radiographic Imaging, 6th ed.
  • Backscatter factors: Grosswendt B. Backscatter factors for X-rays generated at voltages between 10 and 100 kV. Phys Med Biol 1990;35:1233-1245.